Chest
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Guest Access | Sign In via User Name/Password
This Article
Right arrow Full Text (PDF) Free
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Article Archive
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Greenspan, G.
Right arrow Articles by Martin, C.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Greenspan, G.
Right arrow Articles by Martin, C.

Chest, Vol 80, 442-446, Copyright © 1981 by American College of Chest Physicians


ARTICLES

Transcutaneous noninvasive monitoring of carbon dioxide tension

GH Greenspan, AJ Block, LW Haldeman, S Lindsey and CS Martin

We evaluated the accuracy of a new device for continuous noninvasive measurement of cutaneous PCO2. The Hewlett-Packard capnometer (model 47210/HA) works by means of an infrared transducer applied to the forearm over an area of skin that has been stripped of the stratum corneum. Capnometer transcutaneous carbon dioxide pressure (CPCO2) was compared with arterial carbon dioxide pressure (PaCO2) during 60 simultaneously obtained measurements in 13 hemodynamically stable patients. Each patient was studied for 1 1/2 to 5 hours, and a wide range of PaCO2 values (21 to 82 mm Hg) was represented. The data show a clinically significant relationship whereby PaCO2 = CPCO2 - 4.13, with a SE of +/- 2.19 mm Hg. Clinical usefulness of noninvasive cutaneous CO2 monitoring can be foreseen in patients whose ventilatory support is being tapered, in those with respiratory depression caused by various neuromuscular disorders, and in patients with chronic obstructive pulmonary disease and acute respiratory failure. Our results indicate that continuous transcutaneous CPCO2 measurements are safe and accurate and strongly suggest that they can be of clinical usefulness in a select group of hemodynamically stable patients.


This article has been cited by other articles:


Home page
West J Nurs ResHome page
A. Verderber, K. J. Gallagher, and R. Severino
The Effect of Nursing Interventions on Transcutaneous Oxygen and Carbon Dioxide Tensions
West J Nurs Res, February 1, 1995; 17(1): 76 - 90.