Guest Access | Sign In via User Name/Password |
|||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
* From the Department of Internal Medicine II, Divisions of Pneumology and Cardiology (Drs. Arzt, Wensel, Schroll, and Riegger), University of Regensburg, Regensburg; the Center for Pneumology (Drs. Schichtl, Montalvan, Budweiser, and Pfeifer), Donaustauf Hospital, Donaustauf; and Prosper Hospital (Dr. Blumberg), Recklinghausen, Germany.
Correspondence to: Michael Arzt, MD, Department of Internal Medicine II, Pneumology, University of Regensburg, Franz-Josef-Strauβ-Allee 11, 93042 Regensburg, Germany; e-mail: michael.arzt{at}klinik.uni-regensburg.de
Abstract
Background: Treatment with continuous positive airway pressure (CPAP) improves cardiac function in chronic heart failure (CHF) patients with central sleep apnea (CSA)-Cheyne-Stokes respiration (CSR) by stabilizing ventilation, but frequently central apneas and hypopneas persist. Our objective was to test the hypothesis that flow-targeted dynamic bilevel positive airway pressure (BPAP) support (BiPAP autoSV; Respironics; Murrysville, PA) effectively suppresses CSR-CSA in CHF patients.
Methods: We studied 14 CHF patients with CSR-CSA (and residual CSA on positive airway pressure therapy) during 3 consecutive nights: (1) diagnostic polysomnography, (2) CPAP (n = 10) or BPAP (n = 4) titration, and (3) dynamic flow-targeted dynamic BPAP support with an expiratory positive airway pressure (EPAP) set to suppress obstructive respiratory events, and an inspiratory positive airway pressure (IPAP) dynamically ranging between 0 and 15 cm H2O above the EPAP.
Results: CPAP or BPAP significantly reduced the apnea-hypopnea index (AHI) [mean ± SD, 46 ± 4 events/h to 22 ± 4 events/h; p = 0.001] compared to the first night without treatment. Flow-targeted dynamic BPAP support (mean EPAP, 6.5 ± 1.7 cm H2O; maximal IPAP, 21.9 ± 2.1 cm H2O) further reduced the AHI to 4 ± 1/h of sleep compared to the untreated (p < 0.001) and CPAP or BPAP night (p = 0.002). After the first night of flow-targeted dynamic BPAP support, patients rated on an analog scale (range, 0 to 10) the treatment as comfortable (6.9 ± 0.6), and the sleep quality as improved compared to previous nights (7.4 ± 0.6).
Conclusion: Flow-targeted dynamic BPAP support effectively suppresses CSR-CSA in patients with CHF and is well tolerated.
Key Words: cardiology central sleep apnea ventilation
Related Editorial
Chest 2008 134: 4-7.
This article has been cited by other articles:
![]() |
L. K. Brown Filling in the Gaps: The Role of Noninvasive Adaptive Servoventilation for Heart Failure-Related Central Sleep Apnea Chest, July 1, 2008; 134(1): 4 - 7. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |